UV room decontamination reduces C. difficile transmission

Posted

10th July 2015

Research

A study from New York reports a reduction in C. difficile infection associated with the introduction of a pulsed-xenon UV (PX-UV) device. The authors performed a before-after study of the impact of adding PX-UV to the terminal disinfection protocol of rooms housing patients requiring contact precautions, with 12 months before and 12 months after implementation. Overall, there was a non-significant 22% reduction in the rate of CDI (from 1.06 to 0.83 cases per 1000 patient days), and a significant 70% reduction in the ICU (from 1.83 to 0.55 cases per 1000 patient days).

The authors note a change in cleaning contractor towards the end of the pre-intervention period, which could be a confounding factor. And a change in paediatric case mix, which may explain the marginal increase in CDI rates in paediatrics. More importantly, there is a comment in the discussion intimating that PCR testing may have been implemented during the study period, which would be a much more important confounder – but it is not clear whether PCR testing was implemented in the study hospitals or surrounding hospitals.

The authors introduce an unfortunate new term into the literature: “UVD” (ultraviolet light for room decontamination). This is used by the authors as a “catch-all” term to encompass both ultraviolet-C (UVC) and pulsed-xenon UV (PX-UV) devices. This is unfortunate because a recent study shows that a UVC device was, photon for photon, more effective than a PX-UV device.

In the discussion, the authors comment that “Mercury bulbs emit a continuous low pressure light at a single range of 254 nm, and the effect of this UV light is cumulative requiring a longer cycle time of approximately 45 minutes for spore reduction.” Whilst the recommended cycle times for UVC devices are indeed longer than for PX-UV to address contamination with spores, the head-to-head study showed that UVC is more effective than PX-UV when run for the same length of time from the same point in the room. Put another way, a short blast of UVC would do more damage to spores than the same short blast of PX-UV. In the case of this study, if a UVC device had been run as per the PX-UV device, you’d expect to see more of an impact on the spores – and possibly more of an impact on CDI transmission.

The authors also comment that “Pulsed xenon devices also are somewhat safer because mercury-based devices can release toxic gases if they break accidentally.” This issue has been stirred up by manufacturers of PX-UV devices. The amount of mercury in the bulbs is tiny, and there are fail-safe systems in place to ensure that no mercury is released in the event of breakage. The truth is that a fluorescent bulb that you would see in an office block is more hazardous than the bulbs in UVC devices!

So, it seems that in line with other studies, augmenting terminal disinfection with a UV device does reduce C. difficile transmission. But there does not seem to be a good reason to choose PX-UV over a UVC device!

SHARE THIS ARTICLE

Tags

Latest News

Events and Exhibitions

IPC Tour 2024 Highlights

GAMA Healthcare and ACIPC once again joined forces to bring…

Research

Can patient and resident hand hygiene be part of the solution to preventing infections?

The importance of practising proper hand hygiene is widely considered…

Research

Confronting the hidden threat of C. diff

The power of effective surface decontamination The hidden threat of…

Company News

Reflections and learnings from International Infection Prevention Week (IIPW)

International Infection Prevention Week shines a spotlight on the crucial…